Instruction Manual |
1
1.0 Introduction
The Spectrafuge™ 6C is a small benchtop centrifuge designed for separation of various research
samples and is supplied with a 6 x 15 mL rotor. Adapters are available for tubes smaller than 10 mL.
The Spectrafuge 6C reaches speeds of up to 6,500 rpm/4,000 x g.
2.0 Safety Information
Before using the Spectrafuge 6C for the first time, please read this entire manual carefully. If the
equipment is used in a manner not specified by the manufacturer, the protection provided by
the equipment may be impaired.
2.1 Symbols and Conventions
The following chart is an illustrated glossary of the symbols that may be used in this manual or on the
product.
The electrical warning indicates the presence of a potential hazard which could result in
electrical shock.
CAUTION
This symbol refers you to important operating and maintenance (servicing)
instructions within the product Instruction Manual. Failure to heed this information may
present a risk of damage or injury to persons or equipment.
2.2 Operation Safety Precautions
Never use the centrifuge in any manner not specified in these instructions.
Never operate the centrifuge without a rotor properly attached to the shaft.
Never fill tubes while they are in the rotor. Liquid spillage may harm unit.
Never put hands in the rotor area unless the rotor is completely stopped.
Never move the centrifuge while the rotor is spinning.
Never use solvents or flammables near this or other electrical equipment.
Never centrifuge flammable, explosive or corrosive materials.
Never centrifuge hazardous materials outside of a hood or proper containment facility.
Always load the rotor symmetrically. Each tube should be counterbalanced by another tube of the
same type and weight.
Always locate the centrifuge within easy access to an electrical outlet.
Always use only centrifuge tubes designed to withstand centrifugal forces of at least 4,000 x g.
Always use a wrench to tighten rotor nut.
Do not operate the centrifuge if any of the following conditions exist:
The centrifuge has not been installed properly.
The centrifuge is partially dismantled.
Service has been attempted by unauthorized or unqualified personnel.
The rotor has not been installed securely on the motor shaft.
Rotors and accessories not belonging to the standard range are being used without permission being
obtained from the manufacturer to use such rotors and/or accessories in the centrifuge.
Exception: Centrifuge tubes, normally available in the laboratory.
The centrifuge is located in an explosive atmosphere.
Materials to be centrifuged are combustible and/or explosive.
Materials to be centrifuged are chemically reactive.
The rotor load is not properly balanced.