A13914AF
11-4
Utilities
Prime
Priming CC Subsystems
The CC Subsystem option allows independent priming of CC Reagent Delivery Subsystem, CC
Sample Delivery Subsystem and Cuvette Wash.
CAUTION
If any two of the CC Subsystem items are selected, all three will prime
automatically. Check to make certain that all three areas are ready to be primed
if any two are selected.
1
Select:
•
Utils
from the menu bar,
•
1 Prime
,
OR
Type
1
in the
Option Number
field.
•
CC F2
.
2
Select one or a combination of the following check boxes:
Reagent Delivery Subsystem
Sample Delivery Subsystem
Cuvette Wash
.
3
Type the desired number of primes (1–99) in the
Number of primes to repeat
field.
4
Select
Start Prime
. The system status changes to
Priming
and the number of primes remaining
are shown.
5
Upon completion of the requested prime, the number of primes remaining will be blank and the
system status returns to
Standby
.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......