0x01FA002B
4
Hydro Smart Module
Communication
Problem. Stop/Home
may resolve this issue.
CAU:
Hydro SM not
communicating with MSMC.
Check fiber optics
communication loop
Message ID:
0x01FA002B
System stopped. Communication issue
with hydro smart module.
•
Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
the system.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x01FA0034
4
Hydro Illegal Switch
Condition
CAU:
Hydro Illegal Switch
Condition
Message ID:
0x01FA0034
Container:
container
Float Switch sending full and empty
messages at the same time.
•
Unscrew the canister. Wipe the float switch and
make sure the float is not stuck.
•
Replace the Float Switch in the specified
container.
0x025C0001
4
Instrument Computer
Error
CAU:
Debug port abort
Message ID:
0x025C0001
Subsystem:
CAU
Problem Area:
problem area
Likely cause is the Instrument computer
has attempted to access a restricted
area or file.
•
At the Red Message screen, select
•
If the problem continues contact Beckman
Coulter Support Center for additional
information.
0x025C0009
4
Instrument Computer
Error
CAU:
Bus Error
Message ID:
0x025C0009
Subsystem:
CAU
Task Name:
task name
Vector ID:
vector id
Likely cause is the Instrument computer
has received an invalid or unexpected
response. This is not caused by user
input.
•
At the Red Message screen, select
•
If the problem continues contact Beckman
Coulter Support Center for additional
information.
0x0262000F
4
CC Sample Probe
Disabled Due To Level
Sense Errors
CAU:
CC Sample Probe
Disabled Due To Level Sense
Errors
Message ID:
0x0262000F
Subsystem:
RASched
Five consecutive sample level sense
errors – probe, bead, bad alignment or
level sense board problem.
System Response = Pause RA.
•
Make sure Samples are in the correct rack.
•
Perform CC Sample Probe to Cup and CC Sample
Probe Height alignments. Refer to step 3 of the
procedure.
•
Perform CC Sample Probe Level Sense
Alignment (Calibration).
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......