User Defined Reagents
Chemistry Parameters
8
Units
Changing previously defined units for a user defined reagent is not allowed if the chemistry is
included in a current control definition. In addition, changing units invalidates recalled results and
deletes associated reference ranges.
Precision
Precision specifies the number of decimal places for reporting results.
Reaction Direction
A positive reaction direction indicates increasing absorbance with time. A negative reaction
direction indicates decreasing absorbance with time.
Table 8.4
Options for Units
1. mg/dL
11. μg/mL
21. IU/mL
31. negative
41. mg/mL
2. mg/L
12. ng/mL
22. U/mL
32. pg/mL
42. ng/L
3. g/dL
13. μg/dL
23. Rate
33. pg/dL
43. RLU
4. g/L
14. μg/L
24. ng/dL
34. pmol/L
44. S/CO
5. mmol/L
15. nmol/L
25. μIU/mL
35. %Uptake
45. AU/mL
6. μmol/L
16. U/L
26. mIU/mL
36. %GHb
46. AU/L
7. mEq/L
17. other
27. Ku/L
37. %A1c
47. mAU/L
8. nKat/L
18. %
28. nIU/dL
38. GPL
48. mAUL/mL
9. μKat/L
19. mA
29. mIU/L
39. MPL
49. (i) SI
10. IU/L
20. mA/min
30. positive
40. APL
50. % Supp
Table 8.5
Options for Precision
X
X.X
X.XX
X.XXX
Table 8.6
Options for Reaction Direction
Positive
Negative
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......