A13914AF
9-55
Maintenance
Six-Month Maintenance
9
3
Prevent leakage of reagent from the flow cell.
•
Select:
—
Utils
from the menu bar,
—
2 Maintenance
,
—
3 ISE Service
,
—
Continue
.
The flow cell is prepared for electrode maintenance.
4
Refer to the appropriate electrode procedures for servicing other electrodes (
).
•
Chloride Electrode (or
Replace the Chloride Electrode Tip
)
•
CO
2
Measuring Electrode (
•
Sodium Electrode (
Replace the Sodium Measure/Reference Electrode
).
Removal and Installation of the Electrodes
CAUTION
To prevent damage due to electrical static discharge (ESD), wear the wrist ground
strap when you perform this procedure. For further instructions, refer to
in the beginning of this chapter.
1
Disconnect the electrode cable from potassium connector (1) and calcium connector (2).
Figure 9.36
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......