A13914AF
10-10
System Status and Commands
Pause
When to Use Pause
Use
Pause
when the system must be interrupted but there is not an immediate need to stop the
samples in process. Correct uses of
Pause
include:
•
When maintenance is needed but the system is already running.
•
When a rack needs to be removed from the sample carousel but is not needed immediately or
to avoid wasting reagent.
When NOT to Use Pause
Do not use Pause when there is danger involved or when the situation can be handled with another
function.
Pause
1
Select
Instr Cmd
from the menu bar.
2
Select
2 Pause
.
3
To pause the related part of the instrument, from the
Pause
dialog box, select:
•
1 MC Only
,
•
2 CC Only
OR
•
3 Both MC and CC
.
OR
Table 10.3
Alternatives Instead of Pause for Some Situations
Situation...
Instead of Pause...
Imminent danger to a person or the
instrument.
Press the
STOP
button or turn off power.
Reagent needs to be loaded while the
system is running.
Perform a reagent load as usual and the system automatically
interrupts reagent additions to allow the reagent load to continue.
Upon completion of the reagent load, the system will resume
running samples.
Sample on the sample carousel is
needed immediately.
From the
Main
screen select the rack with the sample needed
and select
Unload F2
. Note that any tests that have been started
for the samples in the rack that have not yet had sample added
will be aborted. The rack, however, will be available within
approximately 40 seconds.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......