Maintenance
As-Needed/As-Required Maintenance
9
8
Hold the removed EIC over a sink or absorbent tissue. Using a 20 or 30 mL irrigation syringe
filled with 10% bleach solution, insert the tip of the syringe into the valve port opening (1).
Figure 9.65
9
Flush the 10% bleach solution through the valve port. This flushes the sample inject area. Allow
the bleach solution to sit for one minute.
The one-minute waiting period applies only to the 10% bleach solution, and not to the following
deionized water or 70% isopropyl alcohol.
10
Repeat the flushing procedure through the valve port using deionized water.
11
Repeat the flushing procedure through the valve port using 70% isopropyl alcohol.
12
Repeat the flushing procedure through the valve port using deionized water.