A13914AF
9-44
Maintenance
Three-Month Maintenance
Figure 9.28
MC and CC Sample Syringes (100 μL)
Prepare for Maintenance Mode and Remove the Plunger
Before removing the plunger, the plunger rods must be fully extended to the bottom of the syringe
barrels. Follow the steps below to activate the automated Syringe Rod Replacement procedure.
IMPORTANT
The system status must be in
Standby
to access
Maintenance
mode.
1
Select:
•
Utils
from the menu bar,
•
2 Maintenance
,
•
7 Syringe Rod Replacement
.
2
Select
Continue
to prepare the syringe rod assemblies for maintenance.
OR
Select
Cancel
to immediately return to the Maintenance menu without performing syringe rod
maintenance.
1.
MC Sample Syringe
2.
CC Sample Syringe
3.
Syringe Attached to Luer-lock Fitting
A011873P.EPS
1
2
3
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......