A13914AF
6-2
Sample Programming and Processing
Prior to Programming
WARNING
If your LIS or normal workflow requires the reuse of sample IDs, the sample
programming should be cleared from the DxC at a time interval that is less than
the shortest time of sample ID reuse. The DxC over-write feature is not sufficient
for successfully reusing sample IDs. If this warning is not followed, results from
the new request are merged with tests from incomplete samples that previously
used that ID. In addition, patient demographics at the instrument are from the
previous use of the sample ID. Since demographics are not uploaded from the
instrument, it is possible for these merged results to be uploaded to the LIS which
would contain the current demographics.
Figure 6.1
Program Sample Screen
Prior to Programming
Introduction
Prior to programming, it is necessary to verify rack status and sample carousel status. These
options determine which racks are available for programming, for rack assignment, and for
determining the instrument status.
E0164
8
6S.EPS
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......