A13914AF
4-20
Reagent Load/Calibration
Within-Lot Calibration
2
Select
Options F6
.
3
Select
Within lot calibration
from the pull-down menu.
4
From the pull-down menu, select a desired Configured Chemistry name. Refer to
.
5
Select
Add
to move the chosen Chemistry to the Selected Chemistries area.
6
Repeat Step 5 for other chemistries, if appropriate.
7
Select
OK
to enable the within-lot option for the selected chemistry.
OR
Select
Cancel
to exit without enabling.
Figure 4.11
Within-Lot Calibration Dialog Box
NOTE
If the within-lot calibration mode is disabled, any existing cartridges that are using the within-lot
calibration factor will retain that factor until the cartridge is depleted or until another calibration is
requested. When a chemistry is deselected, the following message appears:
"
If chemistries are deselected, a new calibration factor will need to be generated for all subsequent
reagent packs loaded. Do you want to save changes to within-lot chemistries?
"
E015926S.EPS
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......