A13914AF
6-12
Sample Programming and Processing
Clear Samples
5
A batch may be programmed with Sample IDs only, rack/cup designations only or both, as
follows:
6
To end Sample ID entry and to save the batch programming, select
End F10
.
NOTE
Patient demographics may be entered for a particular Sample ID at any time from the sample
programming screen by typing a Sample ID in the
Sample ID
field and selecting
Demog F2
.
Clear Samples
Introduction
When the test program is processed, if all positions are used, the rack is not available for
programming again until it is cleared. Only the programming information is deleted; the results are
stored and are accessible through the recall function using the appropriate recall option.
Calibrator IDs or rack/position assignments are not cleared from within Sample Programming.
Refer to CHAPTER 4,
Reagent Load/Calibration
,
, for information on calibrator
identification.
IMPORTANT
When clearing data, to verify exactly what will be cleared, first view the load list for the
intended range of samples. The samples recalled on the load list are identical to the samples that will be
cleared.
Identification
Procedure
Rack/cup with Sample IDs
•
Type rack numbers and number of samples desired in
the batch.
•
Select
OK
and type specific Sample IDs.
Sample IDs only
•
Select
Sample IDs Only
.
•
Type specific Sample IDs.
•
Select
Racks F1
to assign specific racks to the IDs or
leave rack/positions empty.
Rack/Pos only
•
Type the rack numbers and number of samples desired
in the batch only. Do not type Sample IDs.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......