A13914AF
9-116
Maintenance
As-Needed/As-Required Maintenance
Replace the CTS Wick
This procedure applies to the UniCel DxC 600/800 and the UniCel DxC 600i Systems with the Closed
Tube Sampling (CTS), 1-Blade Thick Option. The wick is found within the Cap Piercer Assembly.
The Cap Piercer wick and blade are normally replaced every two months during scheduled
maintenance. However, if blood is not continuously wiped from the sample tube caps before they
are loaded onto the instrument, the CTS blade/wick area becomes dirty. This may cause a
contaminated condition. As a result, use this procedure to replace the wick as needed.
CAUTION
All biohazard precautions should be observed when doing maintenance,
service, or troubleshooting activities on the system. This includes but may not be
limited to wearing gloves and eye shields, and washing hands after working on
contaminated portions of the system.
Materials Required:
•
1 medium flat-edge screwdriver
•
1 medium Phillips screwdriver (for DxC 600i systems)
•
1 small tweezers
•
2 plastic trays (approximately 9 inches × 12 inches) (23.9 centimeters × 30.5 centimeters)
•
Absorbant paper towel
•
1 bottle of Auto-Gloss
•
1 Wick Kit (PN A26924)
•
Cotton-tip applicators (6 inches [15.2 centimeters] long, or greater)
•
Disposable pipettes
•
Personal protective gear such as rubber gloves and eye shields
•
Scissors
System Preparation
The Cap Piercer Blade must be lowered before the Wick is replaced. Perform the following
procedure:
1
Select
Utils
from the menu bar.
2
Select
2 Maintenance
.
3
Select
1 CTS Blade Replacement.
Содержание UniCel DxC 600
Страница 4: ...A13914AF iv Revision History ...
Страница 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Страница 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Страница 68: ...A13914AF 1 32 System Description Main Screen and Program Structure Figure 1 20 Program Structure QC Setup ...
Страница 71: ...A13914AF 1 35 System Description Main Screen and Program Structure 1 Figure 1 23 Program Structure Utils ...
Страница 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Страница 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Страница 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Страница 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Страница 407: ......
Страница 490: ...A13914AF Glossary 10 Glossary ...
Страница 497: ......