R&S ESU
Receiver Mode
4.52
Operating Manual 1302.6163.12 - 03
Table 4-1
Default setup of scan table
The diagram parameters to be defined are: start frequency 150 kHz, stop frequency
1 GHz, min. level 0 dBµV, grid range log 100 dB, log. frequency axis and continuous
scan.
The measurement parameters correspond to the settings recommended for over-
view measurements to CISPR 16.
4.5.5.2
Time Domain Scan in the Frequency Domain
Up till now, long EMI measurement times were necessary in EMC test laboratories,
when correct measurements of unknown EUTs had to be made. Time saving proce-
dures as explained in
“Data Reduction and Peak List” on page 4.33
reduced the
total measurement time by reducing the number of quasipeak measurements to a
minimum. But this time is still very long, because it is often in the order of hours,
especially for the CISPR radiated emission test. A way out of this situation can be
time-domain measurements. Whereas in conventional EMI measurement systems,
only the spectrum within the IF filter can be measured during a certain measurement
time of e.g. 100 ms, in TDEMI measurement systems, large parts of the spectrum at
the receiver input can be measured during the same time and it depends on the
dynamic range and the resolution of the A/D conversion system, how accurate the
measurement results are. Systems with higher resolution can provide a higher
dynamic range.
The R&S ESU uses an internal fast A/D converter with up to 80 MHz sampling rate
to convert a spectrum width of up to 7 MHz at a time. The restriction of the FFT width
is due to the built-in preselection.
Range 1
Range 2
Start frequency
150 kHz
30 MHz
Stop frequency
30 MHz
1 GHz
Step width
Auto
Auto
IF bandwidth
9 kHz
120 kHz
Measurement time
1 ms
100 µs
Auto ranging
OFF
OFF
Attenuation mode
Normal
Normal
RF attenuation
10 dB
10 dB
Preamplification
OFF
OFF
This functionality is only available with the R&S ESU-K53 option.
Frequency range
Max. Resolution
Bandwidth
FFT width
<= 150 kHz
50 kHz
150 kHz
150 kHz … 2 MHz
50 kHz
2 MHz
2 MHz … 8 MHz
50 kHz
500 kHz
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...