Operating Manual 1302.6163.12 - 03
6.77
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PPEak:PHOLd:RESult?
This command is used to query the result of the measurement of the positive peak value with active
peak hold function. The query is only possible if the peak hold function has been activated previously
using
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PHOLd
.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of
:MARKer
is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON"
'Switches on the function in screen A.
"CALC:MARK:FUNC:SUMM:PHOL ON"
'Switches on the measurement of the peak value in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:PHOL:RES?"
'Outputs the result of screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-T
This command is only a query and therefore has no *RST value.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS[:STATe]
ON | OFF
This command switches on or off the measurement of the effective (RMS) power in the selected
measurement window. If necessary the function is switched on previously.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of
:MARKer
is
irrelevant. It is only available in the time domain (span = 0).
Example:
"CALC2:MARK:FUNC:SUM:RMS ON"
'Switches on the function in screen B.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:RMS:RESult?
This command queries the result of the measurement of the RMS power value in the selected
measurement window.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of
:MARKer
is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...