Operating Manual 1302.6163.12 - 03
4.321
R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
The permissible setting range is -200 dB to +200 dB in steps of 0.1 dB. Positive off-
sets apply to an amplifier and negative offsets to an attenuator subsequent to the
tracking generator.
The default setting is 0 dB. Offsets <> 0 are marked by the activated enhancement
label
LVL
.
Remote command:
SOUR:POW:OFFS -10dB
4.10.2
Transmission Measurement
This measurement will yield the transmission characteristics of a two-port network.
The built-in tracking generator serves as a signal source. It is connected to the input
connector of the DUT. The input of the R&S ESU is fed from the output of the DUT.
Fig. 4.20 Test setup for transmission measurement
A calibration can be carried out to compensate for the effects of the test setup (e.g.
frequency response of connecting cables).
4.10.2.1
Calibration of Transmission Measurement
SOURCE CAL
The
SOURCE CAL
softkey opens a submenu comprising of the calibration functions
for the transmission and reflection measurement.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...