R&S ESU
Remote Control – Description of Commands
DISPlay Subsystem
6.110
Operating Manual 1302.6163.12 - 03
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:RVALue
<numeric value>
If the tracking generator option or the external generator control option (R&S FSU-B9/R&S FSP-B10)
is available and the normalization in the NETWORK mode is activated, this value defines the power
value assigned to the reference position in the selected measurement window. This value corresponds
to the parameter REFERENCE VALUE in manual operation.
he numeric suffix at TRACe<1...3> is irrelevant.
Example:
"DISP:TRAC:Y:RVAL 0"
'Sets the power value assigned to the reference position to 0 dB (Tracking
Generator/Ext. Generator Control option).
Characteristics:
*RST value:
0 dB (NETWORK mode)
SCPI: device-specific
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1 to 3>:Y[:SCALe]:RPOSition
0 to 100PCT
This command defines the position of the reference value in the selected measurement window. The
numeric suffix in TRACe<1...3> is irrelevant.
In NETWORK mode (Tracking Generator /Ext. Generator Option R&S FSU-B9/R&S FSP-B10) with
active normalization, RPOSition defines the reference point for the output of the normalized
measurement results.
Example:
"DISP:TRAC:Y:RPOS 50PCT"
Characteristics:
*RST value:
100PCT(ANALYZER mode)
50 PCT (NETWORK mode)
SCPI: conform
Mode:
A
DISPlay[:WINDow<1|2>]:TRACe<1...3>:Y[:SCALe]:BOTTom
<numeric_value>
In receiver mode, this command defines the minimum grid level in the current unit for the scan display.
The numeric suffix in TRACE<1...3> is irrelevant.
Example:
"DISP:TRAC:Y:BOTT -20"
'The minimum grid level is set to -20 dBuV (pre-condition: the default unit has not
been changed)
Characteristics:
*RST value: 0
SCPI: conform
Mode:
R
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...