Operating Manual 1302.6163.12 - 03
4.163
R&S ESU
Instrument Functions
Analyzer Mode
CENTER =
MKR FREQ
The
CENTER = MKR FREQ
softkey sets the center frequency to the current marker
or delta marker frequency.
A signal can thus be set to the center of the frequency display range, for example,
so that it can then be examined in detail with a smaller span.
The softkey is not available in the time domain (zero span).
Remote command:
CALC:MARK:FUNC:CENT
Example
A spectrum is displayed with a large span after PRESET. A signal off the center is to
be examined in detail:
REF LEVEL =
MKR LVL
The
REF LEVEL = MKR LVL
softkey sets the reference level to the current marker
level.
Remote command:
CALC:MARK:FUNC:REF
Example
A spectrum is displayed with a large span after PRESET. A signal off the center is to
be examined in detail:
NEXT PEAK
The
NEXT PEAK
softkey sets the active marker/delta marker to the next lower max-
imum of the selected trace.
Remote command:
CALC:MARK:MAX:NEXT
CALC:DELT:MAX:NEXT
[
PRESET
]
R&S ESU is set to the default setting.
[
MKR->
]
MARKER 1 is switched on and automatically jumps to
the largest signal of the trace.
[
CENTER=MKR FREQ
]
The center frequency is set to the marker frequency. The
span is adapted in such a way that the minimum fre-
quency (= 0 Hz) or the maximum frequency is not
exceeded.
[
SPAN
]
The span can, for example, be reduced using the rotary
knob.
[
PRESET
]
R&S ESU is set to the default setting.
[
MKR->
]
MARKER 1 is switched on and automatically jumps to
the largest signal of the trace.
[
CENTER=MKR FREQ
]
The center frequency is set to the marker frequency.
The span is adapted in such a way that the minimum fre-
quency (= 0 Hz) or the maximum frequency is not
exceeded.
[
REF LEVEL = MKR LVL
] The reference level is set to the measured marker level.
[
SPAN
]
The span can, for example, be reduced using the rotary
knob.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...