Operating Manual 1302.6163.12 - 03
4.133
R&S ESU
Instrument Functions
Analyzer Mode
VIEW
The
VIEW
softkey freezes the current contents of the trace memory and displays it.
If a trace is frozen by
VIEW
, the instrument settings can be changed without the dis-
played trace being modified (exception: level display range and reference level, see
below). The fact that the trace and the current instrument setting do not agree any
more is indicated by an enhancement label "*" at the right edge of the grid.
If in the
VIEW
display mode the level display range (
RANGE
) or the reference level
(
REF LEVEL
) are changed, R&S ESU automatically adapts the measured data to
the changed display range. This allows an amplitude zoom to be made after the
measurement in order to show details of the trace.
Remote command:
DISP:WIND:TRAC:MODE VIEW
BLANK
The
BLANK
softkey activates the blanking of the trace on the screen.
Remote command:
DISP:WIND:TRAC OFF
SWEEP
COUNT
The
SWEEP COUNT
softkey activates the entry of the number of sweeps used for
averaging. The allowed range of values is 0 to 30000 and the following should be
observed:
• Sweep Count = 0 means running averaging
• Sweep Count = 1 means no averaging, maxhold or minhold is carried out
• Sweep Count > 1 means averaging over the selected number of sweeps; in the
continuous sweep mode averaging is performed until the set number of sweeps
is attained and is then continued as running averaging.
The default setting is running averaging (Sweep Count = 0). The number of sweeps
used for averaging is the same for all active traces in the selected diagram.
Remote command:
SWE:COUN 64
DETECTOR
See following section
“Selection of Detector” on page 4.137
TRACE MATH
See following section
“Mathematical Functions for Traces” on page 4.142
MIN HOLD
The
MIN HOLD
softkey activates the min peak detector. R&S ESU saves for each
sweep the smallest of the previously stored/currently measured values in the trace
memory. The detector is automatically set to
MIN PEAK
. In this way, the minimum
value of a signal can be determined over several sweeps.
This function is useful e.g. for making an unmodulated carrier in a composite signal
visible. Noise, interference signals or modulated signals are suppressed by the min
hold function whereas a CW signal is recognized by its constant level.
Pressing the
MIN HOLD
softkey again clears the trace memory and restarts the min
hold function.
Remote command:
DISP:WIND:TRAC:MODE MINH
The setting of the sweep count in the
TRACE
menu is equivalent to the setting in
the
SWEEP
menu.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...