Operating Manual 1302.6163.12 - 03
4.131
R&S ESU
Instrument Functions
Analyzer Mode
SELECT
TRACE
The
SELECT TRACE
softkey activates the entry for the active trace (1, 2, 3).
Remote command:
--
(selected via numeric suffix of:
TRACe
)
CLEAR/WRITE
The
CLEAR/WRITE
softkey activates the overwrite mode for the collected mea-
sured values, i.e. the trace is overwritten by each sweep.
In the
CLEAR/WRITE
display mode all the available detectors can be selected. In
the default mode the autopeak detector (setting
AUTO
) is selected.
Each time the
CLEAR/WRITE
softkey is actuated, R&S ESU clears the selected
trace memory and starts the measurement anew.
Remote command:
DISP:WIND:TRAC:MODE WRIT
MAX HOLD
The
MAX HOLD
softkey activates the max peak detector.
The R&S ESU saves the sweep result in the trace memory only if the new value is
greater than the previous one.
The detector is automatically set to
MAX PEAK
. The maximum value of a signal can
thus be determined over several sweeps.
This is especially useful with modulated or impulsive signals. The signal spectrum is
filled up upon each sweep until all signal components are detected in a kind of enve-
lope.
Pressing the
MAX HOLD
softkey again clears the trace memory and restarts the
max hold mode.
Remote command:
DISP:WIND:TRAC:MODE MAXH
AVERAGE
The
AVERAGE
softkey activates the trace averaging function. The average is
formed over several sweeps. Averaging can be performed with any of the detectors
available. If the detector is automatically selected by R&S ESU, the sample detector
is used.
Depending on the setting of AVG MODE, the logarithmic level values or the mea-
sured power/voltage values are averaged.
Averaging is restarted every time the
AVERAGE
softkey is pressed. The trace mem-
ory is always cleared.
Remote command:
DISP:WIND:TRAC:MODE AVER
Side menu
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...