R&S ESU
Remote Control – Description of Commands
Differences in GPIB Behavior between the FSP and the FSE Families of Instruments
6.318
Operating Manual 1302.6163.12 - 03
FSE
[SENSe<1|2>:]AVERage:AUTO
ON | OFF
not available for
R&S ESU
R&S
ESU
FSE
[SENSe<1|2>:]AVERage:COUNt
0 to 32767
FSE
[SENSe<1|2>:]AVERage:COUNt
0 to 32767
R&S
ESU
[SENSe<1|2>:]AVERage:TYPE
VIDeo | LINear
command is used to
select logarithmic or
linear averaging on the
R&S ESU;
therefore parameters are
incompatible to the FSE
FSE
[SENSe<1|2>:]AVERage:TYPE
MAXimum | MINimum |
SCALar
command is used to
select logarithmic or
linear averaging on the
R&S ESU;
therefore parameters are
incompatible to the FSE
R&S
ESU
FSE
[SENSe<1|2>:]AVERage[:STATe<1...3>]
ON | OFF
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:DEMod
<numeric_value>
not available for
R&S ESU and FSET
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:PLL
AUTO | HIGH | MEDium
| LOW
not available for
R&S ESU
R&S
ESU
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo
1Hz to 10MHz
FSET
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo
1Hz to 500MHz
R&S
ESU
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:AUTO
ON | OFF
FSET
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:
EXTernal[:STATe]
ON | OFF
not available for
R&S ESU
R&S
ESU
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio
0.0001 to 1
only numeric values
available. Parameter
ranges differ between
R&S ESU and FSE
FSE
[SENSe<1|2>:]BANDwidth|BWIDth:VIDeo:RATio
0.001 to 1000 | SINe |
PULSe | NOISe
also text parameters are
available. Parameter
ranges differ between
R&S ESU and FSE
not available for FSET
R&S
ESU
FSE
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]
10Hz to 10MHz (anal.
filter) 1Hz to 10MHz
(FFT filter)
FSE: 10Hz to 10MHz
(models 20) 1Hz to
10MHz (models 30)
FSET
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]
10 Hz to 500MHz
R&S
ESU
FSE
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:
AUTO
ON | OFF
FSE
[SENSe<1|2>:]BANDwidth|BWIDth[:RESolution]:
MODE
ANALog | DIGital
not available for
R&S ESU and FSET
Devices
Command
(Sheet 30 of 47)
Parameter
Notes
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...