6.238
Operating Manual 1302.6163.12 - 03
SYSTem:COMMunicate:SERial[:RECeive]:PARity[:TYPE]
EVEN | ODD | NONE
This command defines the parity check for the serial interface (COM).
Parameter:
EVEN: even parity
ODD: odd parity
NONE: no parity check.
Example:
"SYST:COMM:SER:PAR EVEN"
Characteristics:
*RST value: -- (no influence on this parameter, factory default NONE)
SCPI: conform
Mode:
R, A
SYSTem:COMMunicate:SERial[:RECeive]:SBITs
1|2
This command defines the number of stop bits per data word for the serial interface (COM).
Example:
"SYST:COMM:SER:SBITs 2"
Characteristics:
*RST value: -- (no influence on this parameter, factory default 1)
SCPI: conform
Mode:
R, A
SYSTem:COMMunicate:SERial[:RECeive]:PACE
XON | NONE
This command switches on or off the software handshake for the serial interface.
Example:
"SYST:COMM:SER:PACE XON"
Characteristics:
*RST value: -- (no influence on this parameter, factory default NONE)
SCPI: conform
Mode:
R, A
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
This command queries the name of the first printer (in the list of printers) available under Windows NT.
The names of other installed printers can be queried with command
SYSTem:COMMunicate:
PRINter:ENUMerate:NEXT?.
If no printer is configured an empty string is output.
Example:
"SYST:COMM:PRIN:ENUM:FIRS?"
Characteristics:
*RST value: NONE
SCPI: device-specific
Mode:
R, A
SYSTem:COMMunicate:PRINter:ENUMerate:NEXT?
This command queries the name of the next printer installed under Windows NT.
The command
SYSTem:COMMunicate:PRINter:ENUMerate:FIRSt?
should be sent previously
to return to the beginning of the printer list and query the name of the first printer.
The names of other printers can then be queried with NEXT?. After all available printer names have
been output, an empty string enclosed by quotation marks (") is output for the next query. Further
queries are answered by a Query Error.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...