R&S ESU
Instrument Functions
Analyzer Mode
4.218
Operating Manual 1302.6163.12 - 03
In the
SWEEP LIST
table, the individual sweep ranges are set.
Remote command:
SENS:LIST:RANG<1…20>:…
INS BEFORE
RANGE
The
INS BEFORE RANGE
softkey inserts a range in front of the marked line.
Remote command:
--
INS AFTER
RANGE
The
INS AFTER RANGE
softkey inserts a range following the marked line.
Remote command:
--
Range Start:
Start frequency of the range
Range Stop:
Stop frequency of the range
Filter Type:
NORMAL (3dB), EMI (6dB), CHANNEL, RRC, 5-POLE
RBW:
Resolution filter bandwidth
VBW:
Video filter bandwidth
VBW:
Video filter bandwidth; not applicable for CHANNEL and RRC
filters
Sweep Time Mode:
AUTO, MANUAL
Sweep Time:
Sweep time; if
AUTO
is indicated for the sweep time mode, the
automatically calculated sweep time is displayed. If the cell is
edited, the associated sweep time mode is automatically set to
MANUAL.
Detector:
Specifies the range detector:
AUTO PEAK, MAX PEAK, MIN PEAK, SAMPLE, RMS, AVER-
AGE
REF-Level:
Reference level in dBm
The upper edge of the displayed screen area is the value of
the maximum reference level, corrected by the associated
transducer factor.
RF-Attenuator-Mode: AUTO, MANUAL
RF-Attenuator:
Number; as with Sweep Time
PRE-AMP:
ON, OFF
preamplifier selection (options B23, B25, if available)
Sweep Points:
Number of sweep points per range (sweep segment). The
maximum number of sweep points is 8001.
Stop after Sweep:
ON, OFF
if
ON
, the sweep is stopped after the range and continued only
if confirmed by you via a message box (bit 10 of the
OPERation Register” on page 5.24
).
Transd. factor:
NONE or factor (enter via selection list)
Limit check:
ON, OFF (common for all ranges)
Limit:
Limit in dBm (enter via selection list)
A temporary limit line
_SPUL_IN_
is used, based on the range
limits. This temporary limit line is created at measurement start
and may be copied for other purposes.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...