R&S ESU
Remote Control – Basics
Status Reporting System
5.24
Operating Manual 1302.6163.12 - 03
5.7.3.4
STATus:OPERation Register
In the CONDition part, this register contains information on which actions the
instrument is being executing or, in the EVENt part, information on which actions the
instrument has executed since the last reading. It can be read using commands
"STATus:OPERation:CONDition?"
or
"STATus:OPERation[:EVENt]?"
.
5.7.3.5
STATus:QUEStionable Register
This register comprises information about indefinite states which may occur if the
unit is operated without meeting the specifications. It can be queried by commands
STATus:QUEStionable:CONDition?
and
STATus:QUEStionable[:
EVENt]?
.
Bit
No.
Meaning
0
CALibrating
This bit is set as long as the instrument is performing a calibration.
1 to 7
These bits are not used
8
HardCOPy in progress
This bit is set while the instrument is printing a hardcopy.
9
SCAN results available
This bit is set when a block of scan results is available. Must be enabled by TRAC:FEED:
CONT ALWays
10
Sweep Break
This bit is set when end of sweep range is reached (spurious measurement, mode analyzer).
Command “INIT:CONM” has to be used to proceed.
11 to
12
These bits are not used
13
Threshold signal active
14
This bit is not used
15
This bit is always 0
Bit
No.
Meaning
0 to 2
These bits are not used.
3
POWer
This bit is set if a questionable power occurs (cf. also section
4
This bit is not used.
5
FREQuency
The bit is set if a frequency is questionable (cf. section
“STATus:QUEStionable:FREQuency
6
This bit is not used.
7
This bit is not used
8
CALibration
The bit is set if a measurement is performed uncalibrated (equivalent to label "UNCAL")
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...