Operating Manual 1302.6163.12 - 03
5.29
R&S ESU
Remote Control – Basics
Status Reporting System
5.7.3.10
STATus:QUEStionable:POWer Register
This register comprises all information about possible overloads of the unit.
It can be queried with commands
STATus:QUEStionable:POWer:CONDition?
and
STATus:QUEStionable:POWer[:EVENt]?
.
Bit
No.
Meaning
0
OVERload
(Screen A)
This bit is set if the RF input is overloaded. 'OVLD' will then be displayed.
1
UNDerload
(Screen A)
This bit is set if the RF input is underloaded. 'UNLD' will then be displayed.
2
IF_OVerload
(Screen A)
This bit is set if the IF path is overloaded. 'IFOVL' will then be displayed.
3
Overload Trace (Screen A)
This bit is set if the input is overloaded (OVLD or IFOVL) and the Trace Modes AVERAGE ,
MAXHOLD or MINHOLD are active.
'OVTRC' will then be displayed.
This bit only clears if the sweep is started again. A temporary overload will therefore be
detected, e.g. if the overload condition occurs only on sweep number 10 of 1000 during the
average process.
4 to 7
not used
8
OVERload
(Screen B)
This bit is set if the RF input is overloaded. 'OVLD' will then be displayed.
9
UNDerload
(Screen B)
This bit is set if the RF input is underloaded. 'UNLD' will then be displayed.
10
IF_OVerload
(Screen B)
This bit is set if the IF path is overloaded. 'IFOVL' will then be displayed.
11
Overload Trace (Screen B)
This bit is set if the input is overloaded (OVLD or IFOVL) and the Trace Modes AVERAGE ,
MAXHOLD or MINHOLD are active.
'OVTRC' will then be displayed.
This bit only clears if the sweep is started again.
12 to
14
not used
15
This bit is always 0.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...