Operating Manual 1302.6163.12 - 03
4.217
R&S ESU
Instrument Functions
Analyzer Mode
Limit lines are defined and displayed irrespective of the sweep ranges, i.e. they are
not part of the sweep ranges. The unit of the limit lines is restricted to dB or dBm.
The frequency range where measurements are actually performed is set by the start
and stop frequency parameters of the R&S ESU; these parameters are independent
of the sweep ranges. It is thus possible to define sweep ranges for a measurement
task that can be stored and reloaded and to quickly and easily set the frequency
range to be actually measured by means of two parameters; complex editing in the
sweep table is not necessary.
SPURIOUS
ON | OFF
The
SPURIOUS ON/OFF
softkey switches the spurious emissions measurement on
or off according to the current configuration.
Remote command:
SWEEP:MODE LIST
switches the spurious list on
SWEEP:MODE AUTO
switches the spurious list off
SWEEP LIST
The
SWEEP LIST
softkey opens a submenu where predefined sweep ranges can
be edited, or new ranges generated or deleted. A table listing the current sweep
ranges is displayed.
Remote command:
--
EDIT SWEEP
LIST
The
EDIT SWEEP LIST
softkey opens the table for editing the sweep ranges.
When a limit line is defined in steps, the weaker limit is used at the frequency point
with the straight vertical section.
SWEEP LIST
RANGE 1
RANGE 2
RANGE 3
RANGE 4 RANGE 5
Range Start
9 kHz
150 MHz
30 MHz
Range Stop
150 kHz
30 MHz
1 GHz
Filter Type
NORMAL (3dB) NORMAL (3dB) NORMAL (3dB)
RBW
1 kHz
10 kHz
100 kHz
VBW
3 kHz
30 kHz
300 MHz
Sweep time mode
AUTO
AUTO
AUTO
Sweep time
145 ms
300 ms
100 ms
Detector
RMS
RMS
RMS
REF-Level
-10 dBm
-10 dBm
-10 dBm
RF-Att. mode
AUTO
AUTO
AUTO
RF-Attenuator
20 dB
20 dB
20 dB
PRE-AMP
OFF
OFF
OFF
Sweep Points
501
4001
8001
Stop after sweep
OFF
OFF
OFF
Transd. factor
<NONE>
<NONE>
<NONE>
Limit check
ON
ON
ON
Limit
-13 dBm
-13 dBm
-13 dBm
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...