Operating Manual 1302.6163.12 - 03
6.75
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.4.5
CALCulate:MARKer:FUNCtion:SUMMary Subsystem
This subsystem contains the commands for controlling the time domain power functions. These are pro-
vided in the marker subsystem for reasons of compatibility with the FSE family.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary[:STATe]
ON | OFF
This command switches on or off the previously selected time domain power measurements. Thus one
or several measurements can be first selected and then switched on and off together with
CALC:
MARK:FUNC:SUMMary:STATe
.
The function is independent of the marker selection, i.e. the suffix of MARKer is irrelevant. It is only
available in the time domain (span = 0).
Example:
"CALC:MARK:FUNC:SUMM:STAT ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PPEak[:STATe]
ON | OFF
This command switches on or off the measurement of the positive peak value in the selected
measurement window.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of MARKer is
irrelevant. It is only available in the time domain (span = 0).
Example:
"CALC:MARK:FUNC:SUMM:PPE ON"
'Switches on the function in screen A.
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A-T
CALCulate<1|2>:MARKer<1...4>:FUNCtion:SUMMary:PPEak:RESult?
This command is used to query the result of the measurement of the positive peak value in the selected
measurement window. The measurement may have to be switched on previously.
The function is independent of the marker selection, i.e. the numeric suffix <1...4> of MARKer is
irrelevant. It is only available in the time domain (span = 0).
A complete sweep with synchronization to sweep end must be performed between switching on the
function and querying the measured value to obtain a valid query result. This is only possible in single-
sweep mode.
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:MARK:FUNC:SUMM:PPE ON"
'Switches on the function in screen A.
"INIT;*WAI"
'Starts a sweep and waits for the end.
"CALC:MARK:FUNC:SUMM:PPE:RES?"
'Outputs the result of screen A.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...