R&S ESU
Remote Control – Description of Commands
SYSTem Subsystem
6.240
Operating Manual 1302.6163.12 - 03
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
SYSTem:DISPlay:UPDate
ON | OFF
This command switches on or off the update of all display elements during remote control.
Example:
"SYST:DISP:UPD ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
SYSTem:ERRor?
This command queries the earliest entry in the error queue, and deletes it after the readout.
Positive error numbers indicate device-specific errors, negative error numbers are error messages
defined by SCPI (cf. chapter
). If the error queue is empty, the error number 0, "no
error", is returned. This command is identical with the command
STATus:QUEue:NEXT?
.
This command is a query and therefore has no *RST value.
Example:
"SYST:ERR?"
Characteristics:
*RST value: –
SCPI: conform
Mode:
R, A
SYSTem:ERRor:LIST?
This command reads all system messages and returns a list of comma separated strings. Each string
corresponds to an entry in the table SYSTEM MESSAGES.
If the error list is empty, an empty string "" will be returned.
This command is a query and therefore has no *RST value.
Example:
"SYST:ERR:LIST?"
Characteristics:
*RST value: –
SCPI: device-specific
Mode:
R, A
The best performance is obtained when the display output is switched off during
remote control.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...