R&S ESU
Instrument Functions
Tracking Generator – Option R&S FSU-B9
4.330
Operating Manual 1302.6163.12 - 03
4.10.6
External Modulation of the Tracking Generator
MODULATION
The
MODULATION
softkey opens a submenu for selecting different modulation
modes.
The time characteristics of the tracking generator output signal can be influenced by
means of external signals (input voltage range -1 V to +1 V).
Two BNC connectors at the rear panel are available as signal inputs. Their function
changes depending on the selected modulation:
•
TG IN I / AM
and
• TG IN Q / FM
The modulation modes can be combined with each other and with the frequency off-
set function up to a certain degree. The following table shows which modulation
modes are possible at the same time and which ones can be combined with the fre-
quency offset function.
Table 4-7
Simultaneous modes of modulation (tracking generator)
#
= can be combined
EXT AM
The
EXT AM
softkey activates an AM modulation of the tracking generator output
signal.
The modulation signal is applied to the
TG IN I / AM
connector. An input voltage of
1 V corresponds to 100% amplitude modulation.
Switching on an external AM disables the active I/Q modulation.
Remote command:
SOUR:AM:STAT ON
EXT FM
The
EXT FM
softkey activates the FM modulation of the tracking generator output
signal.
The modulation frequency range is 1 kHz to 100 kHz, the deviation can be set in 1-
decade steps in the range of 100 Hz to 10 MHz at an input voltage of 1 V. The phase
deviation h should not exceed the value 100.
Phase deviation h = deviation / modulation frequency
Modulation
Frequency
offset
EXT AM
EXT FM
EXT I/Q
Frequency offset
#
#
#
EXT AM
#
#
EXT FM
#
#
EXT I/Q
#
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...