Operating Manual 1302.6163.12 - 03
6.15
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
CALCulate<1|2>:DELTamarker<1...4>:Y?
This command queries the measured value of the selected delta marker in the indicated measurement
window. The corresponding delta marker will be activated, if necessary. The output is always a relative
value referred to marker 1 or to the reference position (reference fixed active).
To obtain a valid query result, a complete sweep with synchronization to the sweep end must be
performed between the activation of the delta marker and the query of the y value. This is only possible
in single-sweep mode. In receiver mode, the markers can be only activated after a scan is carried out.
Depending on the unit defined with
CALC:UNIT
or on the activated measuring functions, the query
result is output in the units below:
• DBM | DBPW | DBUV | DBMV | DBUA: Output unit DB
• WATT | VOLT | AMPere: Output unit W | V | A
• Statistics function (APD or CCDF) on: Dimensionless output
Example:
"INIT:CONT OFF"
'Switches to single-sweep mode.
"CALC:DELT2 ON"
'Switches on delta marker 2 in screen A.
"INIT;*WAI"
'Starts a sweep and waits for its end.
"CALC:DELT2:Y?"
'Outputs measurement value of delta marker 2 in screen A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:DELTamarker<1...4>:MAXimum[:PEAK]
This command positions the specified delta marker to the current maximum value of the trace. If
necessary, the corresponding delta marker will be activated first.
Example:
"CALC2:DELT3:MAX"
'Sets delta marker 3 in screen B to the maximum value of the associated trace.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
CALCulate<1|2>:DELTamarker<1...4>:MAXimum:NEXT
This command positions the specified delta marker to the next smaller maximum value of the trace.
The corresponding delta marker will be activated first, if necessary.
Example:
"CALC1:DELT2:MAX:NEXT"
'Sets delta marker 2 in screen A to the next smaller maximum value.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...