6.90
Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:PEAKsearch|PSEarch:MARGin
MINimum .. MAXimum
This command defines the margin for the peak search.
The numeric suffix in
CALCULATE<1|2>
is not significant.
Example:
"CALC:PEAK:MARG 5 dB"
Characteristics:
*RST value: 6 dB
SCPI: device-specific
Mode:
A, R
CALCulate<1|2>:PEAKsearch|PSEarch:SUBRanges
1 to 500
This command defines the number of subranges for the peak search.
The numeric suffix in
CALCULATE<1|2>
is not significant.
Example:
"CALC:PEAK:SUBR 10"
Characteristics:
*RST value: 25
SCPI: device-specific
Mode:
A, R
CALCulate<1|2>:PEAKsearch|PSEarch:METHod
SUBRange | PEAK
This command defines the method that is used to determine the level maxima of a scan.
The numeric suffix in
CALCULATE<1|2>
is not significant.
Example:
"CALC:PEAK:METH SUBR"
Characteristics:
*RST value: PEAK
SCPI: device-specific
Mode:
R
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...