R&S ESU
Instrument Functions
Basic Settings
4.298
Operating Manual 1302.6163.12 - 03
The analyzer mode is compatible with the settings of R&S ESU EMI Test Receiver
and R&S FSU Spectrum Analyzer. It facilitates programming of control software for
several instrument types.
Remote command:
SYST:PRES:COMP FSP
OPEN START
MENU
The
OPEN START MENU
softkey opens the windows XP start menu and therefore
provides an easy access to standard windows functions if a mouse is connect.
IF SHIFT
The
IF SHIFT
softkey opens a submenu to activate or deactivate 1
st
IF shifting.
Input signals at a frequency of half the 1
st
IF (in the frequency range of 2270 MHz to
2350 MHz) reduce the dynamic range of the analyzer. This problem only occurs for
low RF attenuation values. It can be overcome by shifting the 1
st
IF.
IF SHIFT OFF
The
IF SHIFT OFF
softkey deactivates the 1
st
IF shifting.
Remote command:
SWE:IF:SHIF OFF
Remote command:
SWE:IF:SHIF OFF
IF SHIFT A
The
IF SHIFT A
softkey is appropriate for input signals in the frequency range of
2270 MHz to 2310 MHz.
Remote command:
SWE:IF:SHIF A
Remote command:
SWE:IF:SHIF A
IF SHIFT B
The
IF SHIFT B
softkey is appropriate for input signals in the frequency range of
2310 MHz to 2350 MHz.
Remote command:
SWE:IF:SHIF B
AUTO
AUTO
automatically selects the suitable 1st IF shifting. The Signal Frequency has to
specified for that purpose in the Signal Frequency dialog.
Remote command:
SWE:IF:SHIF AUTO
SWE:IF:SHIF:FREQ <numeric value>
The 1
st
IF shifting is automatically done for ACP measurements if the center fre-
quency ( = signal frequency) is in the range of 2270 MHz to 2350 MHz. The
IF
SHIFT
setting is therefore ignored for ACP measurements.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...