SENSe:FREQuency Subsystem
The SENSe:FREQuency subsystem defines the frequency axis of the active display. The frequency axis
can either be defined via the start/stop frequency or via the center frequency and span.
The measurement windows are selected by SENSe1 (screen A) and SENSe2 (screen B).
[SENSe<1|2>:]FREQuency:CENTer
0 to f
max
This command defines the receiver frequency or the center frequency in analyzer mode.
Example:
"FREQ:CENT 100MHz"
Characteristics:
*RST value: f
max
/2 with f
max
= maximum frequency
SCPI: conform
Mode:
R, A
[SENSe<1|2>:]FREQuency:CENTer:STEP
0 to f
max
This command defines the step width of the center frequency.
Example:
"FREQ:CENT:STEP 120MHz"
Characteristics:
*RST value: - (AUTO 0.1 × SPAN is switched on)
SCPI: conform
Mode:
R, A
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK
SPAN | RBW | OFF
This command couples the step width of the center frequency to span (span >0) or to the resolution
bandwidth (span = 0) or cancels the couplings.
Parameter:
SPAN = Coupling to frequency display range (for span > 0)
RBW = Coupling to resolution bandwidth (for span = 0)
OFF = manual input, no coupling.
Example:
"FREQ:CENT:STEP:LINK SPAN"
Characteristics:
*RST value: SPAN
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]FREQuency:CENTer:STEP:LINK:FACTor
1 to 100 PCT
This command couples the step width of the center frequency with a factor to the span (span >0) or to
the resolution bandwidth (span = 0).
Example:
"FREQ:CENT:STEP:LINK:FACT 20PCT"
Characteristics:
*RST value: - (AUTO 0.1 × SPAN is switched on)
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...