R&S ESU
Instrument Functions
Analyzer Mode
4.106
Operating Manual 1302.6163.12 - 03
SELECT TRACE
The
SELECT TRACE
softkey selects the trace on which signal tracking is to be per-
formed.
Remote command:
CALC:MARK:FUNC:STR:TRAC 1
4.8.2
Setting the Frequency Span – SPAN Key
The
SPAN
key opens a menu which offers various options for setting the span.
The entry of the span (
SPAN MANUAL
softkey) is automatically active for span > 0
Hz.
For span = 0 Hz the entry for sweep time (
SWEEPTIME MANUAL
) is automatically
active.
With two windows (
SPLIT SCREEN)
displayed at the same time, the input data
always refer to the window selected with the
SCREEN A/B
hotkey.
SPAN
SPAN
MANUAL
The
SPAN MANUAL
softkey activates the window for manually entering the fre-
quency span. The center frequency is kept constant.
Allowed range of span values:
• for the time domain (span = 0): 0 Hz
• and for the frequency domain (span >0): minspan
≤
f
span
≤
f
max
Remote command:
FREQ:SPAN 2GHz
SWEEPTIME
MANUAL
The
SWEEPTIME MANUAL
softkey activates the window for entering the sweep
time manually with Span = 0 Hz. The softkey is not available for Span > 0 Hz.
Remote command:
SWE:TIME 10s
FULL SPAN
The
FULL SPAN
softkey sets the span to the full frequency range of R&S ESU.
Remote command:
FREQ:SPAN:FULL
ZERO SPAN
The
ZERO SPAN
softkey sets the span to 0 Hz. The x-axis becomes the time axis
with the grid lines corresponding to 1/10 of the current sweep time (SWT).
Remote command:
FREQ:SPAN 0Hz
f
span
frequency
span
minspan
smallest selectable span (10 Hz)
f
max
max.
frequency
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...