Operating Manual 1302.6163.12 - 03
4.245
R&S ESU
Instrument Functions
Basic Settings
Remote command:
CALC:LIM3:LOW:OFFS 3dB
CALC:LIM3:UPP:OFFS 3dB
4.9.1.2
Entry and Editing of Limit Lines
A limit line is characterized by
• its name
• the assignment of domain (frequency or time)
• the scaling in absolute or relative times or frequencies
• the vertical unit
• the interpolation
• the vertical scaling
• the vertical threshold (only with relative vertical scaling)
• the margin
• the definition of the limit line as either upper or lower limit.
• the data points for frequency/time and level.
At the time of entry, the R&S ESU immediately checks that all limit lines are in accor-
dance with certain guidelines. These guidelines must be observed if specified oper-
ation is to be guaranteed.
This softkey does not have any effect on limit lines that represent absolute values
for the y-axis.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...