Remote Control – Description of Commands
CALCulate Subsystem
6.26
Operating Manual 1302.6163.12 - 03
CALCulate<1|2>:LIMit<1...8>:NAME
<name of limit line>
This command assigns a name to a limit line. If it does not exist already, a limit line with this name is
created.
The name of the limit line may contain a maximum of 8 characters.
Up to 8 limit lines can be defined at the same time. The command is independent of the measurement
window.
Example:
"CALC:LIM1:NAME 'FM1'"
'Assigns the name 'FM1' to limit line 1.
Characteristics:
*RST value: 'REM1' to 'REM8' for lines 1 to 8
SCPI: device-specific
Mode:
R, A
CALCulate<1|2>:LIMit<1...8>:DELete
This command deletes the selected limit line.
Up to 8 limit lines can be defined at the same time. The command is independent of the measurement
window.
Example:
"CALC:LIM1:DEL"
'Deletes limit line 1.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no *RST value and no query.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...