Operating Manual 1302.6163.12 - 03
4.83
R&S ESU
IF Spectrum Analysis Mode
The measurement results of the active marker (also called
marker values
) are dis-
played in the marker field. The marker info field at the upper right of the display
shows the marker location (here, frequency), the level and the currently selected
trace [T1].
The
MKR
key calls a menu that contains all marker and delta marker standard func-
tions. If no marker is active,
MARKER 1
will be enabled and a peak search on the
trace carried out. Otherwise, the data entry for the marker activated last is opened.
MKR
MARKER
1|2|3|4
MARKER
NORM | DELTA
The
MARKER 1/2/3/4
softkey selects the corresponding marker and activates it.
For further information refer to the descriptions in section
“MARKER 1|2|3|4 MARKER NORM | DELTA” on page 4.64
.
ALL MARKER
OFF
Pressing the
ALL MARKER OFF
softkey switches off all marker.
MKR->TRACE
The
MKR->TRACE
softkey places the marker on a new trace.
For further information refer to the descriptions in section
LINK MKR1
AND DELTA1
With the softkey
LINK MKR1 AND DELTA1
the delta marker1 can be linked to
marker1, so if the x-axis value of the marker1 is changed the delta marker1 will fol-
low on the same x-position. The link is default off, and can be switched on.
Example for setup:
1.
PRESET
2.
TRACE
|
MAX HOLD
3.
TRACE
|
SELECT TRACE
|
2
|
AVERAGE
4.
MKR
(Switches marker1 on)
5.
MARKER NORM DELTA
|
DELTA
(Switches delta marker 1 on)
6.
MKR->
|
MKR->TRACE
|
2
7.
LINK MKR1 AND DELTA1
Now select the marker1 (by switching
MARKER1
from
DELTA
back to
NORM
) and
when changing the x-axis value (by rotary knob or
UP/DOWN
keys), the delta
marker1 will follow automatically.
MARKER 1|2|3|4 MARKER
NORM | DELTA
Side menu
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...