Operating Manual 1302.6163.12 - 03
5.5
R&S ESU
Remote Control – Basics
Starting Remote Control
During program execution it is recommended to activate the display of results by
means of
"SYSTem:DISPlay:UPDate ON"
so that it is possible to follow the
changes in the device settings and the recorded measurement curves on the
screen.
5.4.2
Remote Control via GPIB
5.4.2.1
Setting the Device Address
In order to operate the instrument via the GPIB, it must be addressed using the set
GPIB address. The GPIB address of the instrument is factory-set to 20. It can be
changed manually in the
SETUP - GENERAL SETUP
menu or via remote control.
Addresses 0 to 30 are permissible.
Manually:
1. Call
SETUP - GENERAL SETUP
menu
2. Enter desired address in table
GPIB-ADDRESS
3. Terminate input using the
ENTER
key
Via GPIB:
CALL IBFIND("DEV1", analyzer%)
'Open port to the instrument
CALL IBPAD(analyzer%, 20)
'Inform controller about old address
CALL IBWRT(analyzer%, "SYST:COMM:GPIB:ADDR 18")
'Set instrument to new address
CALL IBPAD(analyzer%, 18)
'Inform controller about new address
5.4.2.2
Return to Manual Operation
Return to manual operation is possible via the front panel or remote control.
Manually:
Press the
LOCAL
softkey or the
PRESET
key
If the instrument is exclusively operated in remote control, it is recommended to
switch on the power-save mode (
POWER SAVE
). In this mode, the required dis-
play is completely switched off after a preset time.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...