R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.186
Operating Manual 1302.6163.12 - 03
"SENSe:LIST:POWer
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs a measurement sequence with the following settings:
"SENSe:LIST:POWer?
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,CFIL,30kHz,100kHz,434us,0"
performs the same measurement and returns the result list immediately after the last frequency point.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
Step
Freq.
[MHz]
Ref Level
RF Att
el Att
Filter type
RBW
VBW
Meas Time
TRG Level
(reserved)
1
935.2
-20 dBm
10 dB
OFF
Normal
1 MHz
3 MHz
434 us
0
2
935.4
-20 dBm
10 dB
OFF
Channel
30 kHz
100 kHz
434 us
0
3
935.6
-20 dBm
10 dB
OFF
Channel
30 kHz
100 kHz
434 us
0
• The measurement is performed in the time domain and therefore the span is set
to 0 Hz. If the time domain is abandoned, the function is automatically switched
off.
• The measurement is not compatible with other measurements, especially as far
as marker, adjacent channel power measurement or statistics are concerned.
The corresponding commands thus automatically deactivate the function.
• The function is only available in remote control. It is deactivated when switching
the instrument back to LOCAL.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...