Operating Manual 1302.6163.12 - 03
6.107
R&S ESU
Remote Control – Description of Commands
DISPlay Subsystem
DISPlay[:WINDow<1|2>]:ACTive?
This command returns the active measurement window. The numeric response has following
meaning:
Example:
"DISP:WIND:ACT?"
returns the active window
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
R, A
The numeric suffix at WINDow<1|2> is irrelevant.
DISPlay[:WINDow<1|2>]:SELect
This command selects the active measurement window. WINDow1 corresponds to SCREEN A,
WINDow2 to SCREEN B.
In full screen mode, the measurements are only performed in the active measurement window.
Measurements are therefore initiated in the active window and result queries (marker, trace data and
other results) answered also in the active window.
Initiating measurements and querying results in the inactive window yields an error message
(execution error).
In split screen mode, the selection of the active window for result queries is irrelevant.
In full screen mode, settings
can also be performed in the inactive measurement window. They
become effective as soon as the corresponding window becomes active.
Example:
"DISP:WIND2:SEL"
'Selects screen B as active measurement window.
Characteristics:
*RST value: SCREEN A active
SCPI: device-specific
Mode:
R, A
This command is an event and therefore has no query.
DISPlay[:WINDow<1|2>]:SIZE
LARGe | SMALl
This command switches the measurement window for channel and adjacent channel power
measurements to full screen or half screen. Only "1" is allowed as a numerical suffix.
Example:
"DISP:WIND1:SIZE LARG"
'Switches the measurement window to full screen.
Characteristics:
*RST value: SMALl
SCPI: device-specific
Mode:
A
1
Screen A
2
Screen B
3
Screen C
4
Screen D
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...