Operating Manual 1302.6163.12 - 03
6.197
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]MPOWer:RESult:MIN?
This command queries the minimum power value in a multiple burst power measurement as
configured and initiated with
SENSe:MPOWer[:SEQuence]
. The unit used for the return values is
always dBm.
If no measurement result is available, the command will return a query error.
Example:
*ESE 1
*SRE 32
'Configuration of status reporting systems for the generation of an SRQ on
operation complete
SENSe:MPOWer
935.2MHz,1MHz,434us,VIDEO,50PCT,5us,MEAN,20;*OPC
'Configuring and starting the measurement
...
'Further actions of the control computer during measurement
On SRQ:
'Response to service request
SENSe:MPOWer:RESult:MIN?
Characteristics:
*RST value: --
SCPI: instrument-specific
Mode:
A-F, A-T
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...