R&S ESU
Instrument Functions
Basic Settings
4.276
Operating Manual 1302.6163.12 - 03
The table with the data of the marked set (the
EDIT
softkey) or an empty table in
which the following entries are preset (the
NEW
softkey) is displayed:
The characteristics of the set can be entered in the header field of the table, the sub-
ranges in the columns of the set.
An overwritten transducer set remains stored in the background as long as the
edited factor is stored with the
SAVE TRD SET
softkey or until the table is closed. A
set overwritten by mistake can be restored by leaving the entry.
Name
- Entry of name
A maximum of 8 characters is permissible for the name. The characters have to
comply with the convention of DOS file names. If the name exceeds 8 characters,
the name is truncated. The unit automatically stores all transducer sets with the
extension .TDS.
If an existing name is changed, the set stored under the previous name is retained
and will not be overwritten automatically with the new name. The previous set can
be deleted at a later time using
DELETE FACTOR/SET
. Thus, sets can be copied.
Remote command:
CORR:TSET:SEL <name>
Unit
- Selection of unit
The unit of the transducer set is selected from a selection box activated by the
ENTER key.
The unit should be selected prior to the entry as it determines the settable trans-
ducer factors. The preset unit for new sets is "dB".
Unit:
dB
Break:
OFF
Name
Entry of the name
Unit
Selection
of
unit
Break
Activating the query when changing the subrange
Comment
Entry of a comment
Start
Entry of the start frequency of the subrange
Stop
Entry of the stop frequency of the subrange
Factors
Selection of the transducer factors for the subrange
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...