R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.202
Operating Manual 1302.6163.12 - 03
[SENSe<1|2>:]POWer:ACHannel:FILTer:STATe:CHANnel<1...12>
ON | OFF
This command switches the weigthing filter state for the TX channel on or off. It is only available if Fast
ACP is not active and no predefine standard is selected.
This command changes the setting for the TX channel. The numeric suffix at CHANnel is ignored.
Example:
"POW:ACH:FILT:STAT:CHAN ON"
'activates the weighting filter
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]POWer:ACHannel:FILTer:STATe:ACHannel
ON | OFF
This command switches the weigthing filter state for the adjacent channel on or off. It is only available
if Fast ACP is not active and no predefine standard is selected.
This command changes the setting for the alternate channels as well.
Example:
"POW:ACH:FILT:STAT:ACH ON"
’activates the weighting filter
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]POWer:ACHannel:FILTer:STATe:ALTernate<1 to 11>
ON | OFF
This command switches the weigthing filter state for the alternate channel on or off. It is only available
if Fast ACP is not active and no predefine standard is selected.
This command changes the setting for all alternate channels and for the adjacent channel as well. The
numeric suffixes at ALTernate are irrelevant for this command.
Example:
"POW:ACH:FILT:STAT:ALT ON"
'activates weighting filter
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]POWer:ACHannel:MODE
ABSolute | RELative
This command toggles between absolute and relative adjacent channel measurement.
For the relative measurement the reference value is set to the currently measured channel power by
command
SENSe:POWer:ACHannel:REFerence:AUTO ONCE
.
The command is only available in the frequency domain (span > 0).
Example:
"POW:ACH:MODE REL"
Characteristics:
*RST value: ABSolute
SCPI: device-specific
Mode:
A-F
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...