Operating Manual 1302.6163.12 - 03
6.65
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.4.3
CALCulate:MARKer:FUNCtion:POWer Subsystem
The CALCulate:MARKer:FUNCtion:POWER subsystem contains the commands for control of power
measurement.
CALCulate<1|2>:MARKer<1...4>:FUNCtion:POWer:SELect
ACPower | CPOWer | MCACpower |
OBANdwidth | OBWidth | CN | CN0
This command selects – and switches on – one of the above types of power measurement in the
selected measurement window. This function is independent of the selected marker, i.e. the numerical
suffix <1...4> appended to MARKer has no effect.
The channel spacings and channel bandwidths are configured in the
SENSe:POWer:ACHannel
subsystem.
Please note the following:
If CPOWer is selected, the number of adjacent channels (command:
[SENSe:]POWer:ACHannel:
ACPairs)
is set to 0. If ACPower is selected, the number of adjacent channels is set to 1, unless
adjacent channel power measurement is switched on already.
This command is a query and therefore has no *RST value.
The occupied bandwidth measurement is performed for the trace on which marker 1 is positioned. To
select another trace for the measurement, marker 1 is to be positioned on the desired trace by means
of
CALC:MARK:TRAC 1|2|3
.
Parameter:
ACPower: adjacent channel power measurement with a single carrier signal
CPOWer: Channel power measurement with a single carrier signal (equivalent to
adjacent channel power measurement with
NO. OF ADJ CHAN
= 0)
MCACpower: Channel/adjacent channel power measurement with several carrier
signals
OBANdwidth | OBWidth: Measurement of occupied bandwidth
CN: Measurement of carrier-to-noise ratio
CN0: Measurement of carrier-to-noise ratio referenced to 1 Hz bandwidth
Example:
"CALC:MARK:FUNC:POW:SEL ACP"
'Switches on adjacent channel power measurement in window A.
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
A-F
The channel/adjacent channel power measurement is performed for the trace
selected with
SENSe:POWer:TRACe 1|2|3
.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...