R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.184
Operating Manual 1302.6163.12 - 03
Example:
'Configuration of the status reporting system for the generation of an SRQ
on operation complete
*ESE 1
*SRE 32
'Configuring and starting the measurement
"SENSe:LIST:POWer
935.2MHz,-20dBm,10dB,OFF,NORM,1MHz,3MHz,434us,0,
935.4MHz,-20dBm,10dB,OFF,NORM,30kHz,100kHz,434us,0,
935.6MHz,-20dBm,10dB,OFF,NORM,30kHz,100kHz,434us,0;
*OPC"
'Further actions of the control computer during measurement
to
'Response to service request
On SRQ:
SENSe:LIST:POWer:RESult?
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]LIST:POWer[:SEQuence]
<analyzer freq>,<ref level>,<rf att>,OFF, <el att>,
<filter type>,<rbw>,<vbw>, <meas time>,<trigger level>
This command configures the list of settings (max. 100 entries) for the multiple power measurement
and starts a measurement sequence. When synchronizing the command with *OPC, a service request
is generated as soon as all frequency points are processed and the defined number of individual
measurements is reached.
To reduce the setting time, all indicated parameters are set up simultaneously at each test point.
The query form of the command processes the list and immediately returns the list of results. The
number of results per test point depends on the setting of the "
SENSe:LIST:POWer:SET
" command.
Parameter:
The following parameters are the settings for an individual frequency point. They
are repeated for every other frequency point.
For reasons of compatibility with the FSP instrument family, input OFF is required
after parameter <rf att>.
<analyzer freq>:
Receive frequency for the signal to be measured (= center frequency in manual
operation)
Range of values:
0 Hz to max. frequency, depending on the instrument
model.
<ref level>:
Reference level
Range of values:
+30 dBm to -75 dBm in 5 dB steps
<rf att>:
RF input attenuation
Range of values:
0 dB to 75 dB in 5 dB steps
OFF:
always OFF
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...