Operating Manual 1302.6163.12 - 03
6.223
R&S ESU
Remote Control – Description of Commands
SOURce Subsystem
6.17.2
SOURce:EXTernal Subsystem
The SOURce:EXTernal subsystem controls the operation of the unit with option Ext. Generator Control
(R&S FSP-B10). The commands are only valid for the selected window, with SOURce1 changing the
setting in screen A and SOURce2 the setting in screen B.
The selection of the external generator 1 or 2 is via EXTernal<1|2>.
SOURce<1|2>:EXTernal<1|2>[:STATe]
ON | OFF
This command activates or deactivates the external generator selected with
SOUR:EXT<1|2>:FREQ:
SWE ON
in the selected window.
The suffix behind EXTernal is irrelevant for this command.
This command is only available in connection with option Ext. Generator Control R&S FSP-B10.
Example:
"SYST:COMM:RDEV:GEN1:TYPE 'SMP02'"
'Selects SMP02 as generator 1.
"SYST:COMM:RDEV:GEN1:LINK TTL"
'Selects GPIB + TTL link as interface.
"SYST:COMM:RDEV:GEN1:ADDR 28"
'Sets the generator address to 28.
"
SOUR:EXT1:FREQ:SWE ON"
'Activates the frequency sweep for generator 1.
"SOUR:EXT ON"
'Activates the external generator
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
SOURce<1|2>:EXTernal<1|2>:FREQuency[:FACTor]:DENominator
<numeric_value>
This command defines the denominator of the factor with which the analyzer frequency is multiplied in
order to obtain the transmit frequency of the selected generator 1 or 2 in the selected window.
The commands of the SOURce:EXTernal subsystem assume that the addressed
generator was correctly configured with the commands of subsystem
SYSTem:
COMMunicate:GPIB:RDEVice:GENerator
.
If no external generator is selected, if the GPIB address is not correct or the gener-
ator is not ready for operation, an execution error will be generated.
Select the multiplication factor in a way that the frequency range of the generator
is not exceeded by the following formula:
if applied to the start and stop frequency of the analyzer.
This command is only valid in combination with option Ext. Generator Control
R&S FSP-B10.
F
Generator
F
Analyzer
Numerator
Denominator
-------------------------------
⋅
F
Offset
+
=
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...