Operating Manual 1302.6163.12 - 03
6.287
R&S ESU
Remote Control – Description of Commands
GPIB Commands of HP Models 856xE, 8566A/B, 8568A/B and 8594E
6.23.6
Model-Dependent Default Settings
If the GPIB language is switched over to an 85xx model, the GPIB address will automatically be switched
over to 18 provided that the default address of the R&S ESU (20) is still set. If a different value is set, this
value will be maintained. Upon return to SCPI, this address will remain unchanged.
The following table shows the default settings obtained after a change of the GPIB language and for the
commands IP, KST and RESET:
6.23.7
Data Output Formats
In the case of the SCPI and IEEE488.2 standards, the output formats for numerical data are flexible to a
large extent. The output format for the HP units, by contrast, is accurately defined with respect to the
number of digits. The memory areas for reading instrument data have therefore been adapted accord-
ingly in the remote-control programs for instruments of this series.
Therefore, in response to a query, the R&S ESU returns data of the same structure as that used by the
original instruments; this applies in particular to the number of characters returned.
Two formats are currently supported when trace data is output: Display Units (command O1) and physi-
cal values (command O2, O3 or TDF P). As to the "Display Units" format, the level data of the R&S ESU
is converted to match the value range and the resolution of the 8566/8568 series. Upon transition to the
REMOTE
state, the R&S ESU is reconfigured such that the number of test points (trace points) corre-
sponds to that of the 85xx families (1001 for 8566A/B and 8568A/B, 601 for 8560E to 8565E, 401 for
8594E).
Model
# of Trace Points
Start Freq.
Stop Freq.
Ref Level
Input Coupling
8566A/B
1001
2 GHz
22 GHz
0 dBm
DC
8568A/B
1001
0 Hz
1.5 GHz
0 dBm
AC
8560E
601
0 Hz
2.9 GHz
0 dBm
AC
8561E
601
0 Hz
6.5 GHz
0 dBm
AC
8562E
601
0 Hz
13.2 GHz
0 dBm
AC
8563E
601
0 Hz
26.5 GHz
0 dBm
AC
8564E
601
0 Hz
40 GHz
0 dBm
AC
8565E
601
0 Hz
50 GHz
0 dBm
AC
8594E
401
0 Hz
3 GHz
0 dBm
AC
Notes regarding the set stop frequency:
• The stop frequency given in the table may be limited to the corresponding
frequency range of the R&S FSUP.
• Command LF sets the stop frequency for 8566A/B to a maximum value of 2
GHz.
Note regarding the number of test points (trace points):
• The number of trace points is switched over only upon transition to the REMOTE
state.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...