Operating Manual 1302.6163.12 - 03
4.63
R&S ESU
Receiver Mode
CONT AT HOLD
With the
CONT AT HOLD
softkey the scan is continued where it was interrupted.
The scan is always continued with the settings in the scan table.
Remote command:
--
STOP SCAN
The
STOP SCAN
softkey stops the scan. Upon restart, scanning starts at the begin-
ning. The results of the performed measurements are lost.
Remote command:
:ABOR
4.5.6
Triggering the Scan – TRIG
Key
The
TRIG
key opens a menu for selection of the trigger sources and the trigger
polarity. The active trigger mode is indicated by highlighting the corresponding soft-
key.
To indicate that a trigger mode other than
FREE RUN
has been set, the enhance-
ment label
TRG
is displayed on the screen. If two windows are displayed, TRG
appears next to the appropriate window.
TRIG
FREE RUN
The
FREE RUN
softkey activates the free-run sweep mode, i.e. start of a scan is not
triggered. Once a measurement is completed, another is started immediately.
FREE RUN
is the default setting of R&S ESU.
Remote command:
TRIG:SOUR IMM
VIDEO
The
VIDEO
softkey activates triggering via the displayed voltage.
For the video triggering mode, a level line showing the trigger threshold is displayed.
Using the level line, the threshold can be adjusted between 0% and100% of the dia-
gram height.
Remote command:
TRIG:SOUR VID
TRIG:LEV:VID 50 PCT
EXTERN
The
EXTERN
softkey activates triggering via a TTL signal at the input connector
EXT TRIGGER/GATE
on the rear panel.
Remote command:
TRIG:SOUR EXT
POLARITY
POS | NEG
The
POLARITY POS/NEG
softkey selects the polarity of the trigger source.
The scan starts after a positive or negative edge of the trigger signal. The selected
setting is highlighted.
The selection is valid for all trigger modes with the exception of
FREE RUN
.
The default setting is
POLARITY POS.
Remote command:
TRIG:SLOP POS
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...