R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.174
Operating Manual 1302.6163.12 - 03
Characteristics:
*RST value: depends on range
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:RELative:STOP
<numeric_value>
This command defines the relative limit at the stop frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBc
Example:
"ESP:RANG5:LIM:REL:STOP -40"
'Relative limit at stop frequency in range 5 is -40 dBc
Characteristics:
*RST value: depends on range
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:STATe
ABSolute | RELative | AND | OR
This command sets the type of limit check for all ranges of Spectrum Emission Mask measurement.
• ABSolute
Checks only the absolute limits defined
• RELative:
Checks only the relative limits. Relative limits are defined as relative to the measured power in the
reference range.
• AND
Combines the absolute and relative limit. The limit check fails when both limits are violated.
• OR
Combines the absolute and relative limit. The limit check fails when one of the limits is violated.
The numeric suffixes at SENSe<1|2> and the numeric suffix at RANGe<1..20> are irrelevant.
Example:
"ESP:RANG5:LIM -40"
'limit in range 5 is -40 dBc
"ESP:RANG7:LIM -20"
'limit in range 7 is -20 dBc
"ESP:RANG:LIM:STAT ABS"
'activates relative limit check
Characteristics:
*RST value: Limit check is OFF
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:RLEVel
<numeric_value>
This command defines the reference level of a range in the Spectrum Emission Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:RLEV -30"
'sets the reference level in range 2 to -30 dBm
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...