R&S ESU
Receiver Mode
4.64
Operating Manual 1302.6163.12 - 03
4.5.7
Marker Functions – MKR Key
MKR
MARKER
1|2|3|4
MARKER
NORM | DELTA
The
MARKER 1|2|3|4
softkey selects the corresponding marker and activates it.
MARKER 1
is always the normal marker. After they have been switched on,
MARK-
ERS
2 to 4
are delta markers that refer to
MARKER 1
. These markers can be con-
verted into markers with absolute value display by means of the
MARKER NORM
DELTA
softkey. When
MARKER 1
is the active marker, pressing the
MARKER
NORM DELTA
softkey switches on an additional delta marker.
Pressing the
MARKER 1/2/3/4
softkey again switches off the selected marker.
Example:
Remote command:
CALC:MARK ON;
CALC:MARK:X <value>;
CALC:MARK:Y?
CALC:DELT ON;
CALC:DELT:MODE ABS|REL
CALC:DELT:X <value>;
CALC:DELT:X:REL?
CALC:DELT:Y?
When several traces are being displayed, the marker is set to the maximum value
(peak) of the active trace which has the lowest number (1 to 3). In case a marker is
already located there, it will be set to the frequency of the next lowest level (next
peak).
MARKER 1|2|3|4
MARKER NORM | DELTA
[
PRESET
]
R&S ESU is set to the default setting.
[
MKR
]
On calling the menu,
MARKER 1
is switched on
('1' highlighted in the softkey) and positioned on
the maximum value of the trace. It is a normal
marker and the
MARKER NORMAL
softkey is
highlighted.
[
MARKER 2
]
MARKER 2
is switched on ('2' highlighted in the
softkey). It is automatically defined as a delta
marker on switching on so the
DELTA
is high-
lighted on softkey
MARKER NORM | DELTA
. The
frequency and level of
MARKER 2
with reference
to
MARKER 1
are output in the marker info field.
[
MARKER NORM | DELTA
] The
MARKER NORM | DELTA
softkey is high-
lighted.
MARKER 2
becomes a normal marker.
The frequency and level of
MARKER 2
are output
as absolute values in the marker info field.
[
MARKER 2
]
MARKER 2
is switched off.
MARKER 1
is the
active marker for entry. The frequency and level
of
MARKER 1
are output in the marker info field.
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...