R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
6.172
Operating Manual 1302.6163.12 - 03
[SENSe<1|2>:]ESPectrum:RANGe<1…20>[:FREQuency]:STOP
<numeric_value>
This command sets the relative stop frequency of a range in the Spectrum Emission Mask
measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:STOP 2.52MHZ"
'stop frequency of range 2 to +2.52MHz
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:INPut:ATTenuation
<numeric_value>
This command defines the RF attenuation of a range in the Spectrum Emission Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"
ESP:RANG2:INP:ATT 30db"
'RF attenuation of range 2 to 30 dB
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:INPut:ATTenuation:AUTO
ON | OFF
This command switches the attenuation mode of a range in the Spectrum Emission Mask
measurement to AUTO (ON) or MANUAL (OFF).
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:INP:ATT:AUTO ON"
'attenuation mode is AUTO for range 2
Characteristics:
*RST value: ON
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1…20>:INPut:GAIN:STATe
ON | OFF
This command switches the preamplifier of a range in the Spectrum Emission Mask measurement on
or off.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG2:INP:GAIN:STAT ON"
' activates the preamplifier for range 2
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
A
Summary of Contents for 1129.7246.03
Page 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual ...
Page 25: ......
Page 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03 ...
Page 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03 ...
Page 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03 ...
Page 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03 ...
Page 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03 ...
Page 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03 ...
Page 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03 ...
Page 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU ...
Page 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03 ...